AXIOMET AX-EL300W60A Расширенная инструкция онлайн [8/72] 527990

AXIOMET AX-EL300W Расширенная инструкция онлайн [8/72] 527991
/// MIN /// Parameter minimum value
/// MAX /// Parameter maximum value
/// DELAY /// Test time
Note:
- The parameters of Config, Protection, Trigger Set and Auto Test in main menu will be saved automatically
after been modified. The parameter saved in location 0 will be recalled automatically every time when the
load is turned on.
- The capital letters of the menu will be shown on the LCD. The numbers of shown characters should not
exceed 6.
2.7.2. Mode and Parameter Setting Menu
The electronic load can be excuted in the following modes.
1: Constant Current mode (CC)
2: Constant Voltage mode (CV)
3: Constant Resistance mode (CR)
4: Constant Power mode (CP)
In the basic mode, press corresponding key to enter into mode and parameter setting menu, which is shown
as below:
Key name /// Function and parameter /// Description
I-SET /// SCCL 01.000A /// Constant current mode: immediate current level
V-SET /// SCVL 80.00V /// Constant voltage mode: immediate voltage level
R-SET /// SVHCRL 02.000R /// Constant resistance mode: immediate resistance level
P-SET /// SCPV 10.000W /// Constant power mode: immediate power level
2.7.3. Transient Operation Menu
In the basic mode, turn off the input, press “Tran key in desired mode to enable its transient operation
and enter into parameters setting menu.
Function /// Description
STATE /// Transient operation status and enable the function
MODE /// Transient operation mode; Continuous: 0 COUNT; Pulsed: 1 PULSE; Toggled: 2 TOGGLE
LEVELL /// Transient low level
LEVELH /// Transient high level
TIME L /// Time for transient low level (0-99999ms), unit is ms
TIME H /// Time for transient high level (0-99999ms), unit is ms
TIME R /// Time for transient rising edge (0.25-99999ms) , unit is ms
TIME F /// Time for transient falling edge (0.25-99999ms) , unit is ms
Note:
Transient operation may be used in CC, CV, and CR modes.