Fluke 1760 [55/110] Apparent power

Fluke 1760 [55/110] Apparent power
Operation
Methods of Measurement/Formulas 3
3-15
where
i
U rms value of the voltage bin with the frequency i5 Hz
i
I rms value of the current bin with the frequency i5 Hz
i
ϕ
phase angle difference between voltage and current of the frequency i5 Hz
The 3-phase real power sum is calculated as
321 LLLsum
PPPP ++=
Apparent power
Apparent power
S
is calculated by multiplying rms values of voltage and current of the
given aggregation.
IUS =
The 3-phase apparent power sum is calculated as
321 LLLsum
SSSS ++=
Reactive power
The harmonic reactive power
h
Q is also derived from the results of the FFT calculations
of voltage and current. The index h in
h
Q indicates that the calculated reactive power
component is derived from the harmonics (FFT) calculation.
=
=
1023
0
sin
i
iiih
IUQ
ϕ
Because every frequency component of the calculation has a sign,
h
Q is also signed. All
other reactive power components are unsigned (always positive) by definition.
The 3-phase harmonic reactive power sum is calculated as
321 LhLhLhsumh
QQQQ ++=
The total reactive power
tot
Q is calculated by
22
PSQ
tot
=
The 3-phase total reactive power sum is calculated as
321 LtotLtotLtotsumtot
QQQQ ++=
The remaining reactive power component
d
Q (d stands for distortion) contains the rest of
the reactive power phenomena (mainly distortion reactive power and modulation reactive