Fluke 1760 [92/110] Current clamp 20 a 200 a ac

Fluke 1760 [92/110] Current clamp 20 a 200 a ac
1760
Users Manual
6-14
a680501048-phase-50a.bmp
Dimensions: See 2540445.
Current Clamp 20 A/200 A AC
This current probe has been designed for non intrusive, accurate measurements of small
AC currents. Using the latest technologies (internal memory for calibration data) provides
current ranges from 0.2 A up to 200 A. The measurement range can be selected in the PQ
Analyze software: IAC20 or IAC200.
Electrical Characteristics
Nominal current In: 20 A, 200 A AC rms
Measuring ranges: 0.2 A – 20 A or 2 A – 200 A
Crest factor: < 3
Peak current: 74 A, 300 A
Max. non destructive current: Up to 300 A rms
Conductor position influence: < 0.5 % of range for 50/60 Hz
Error due to adjacent conductor: 15 mA/A for 50 Hz
Phase error (to reference conditions): < ±0.5 degrees
Frequency (clamp without the
instrument):
40 Hz – 10 kHz (3 dB)
Temperature coefficient: 0.015 % of range/ °C
Safety: 300 V CAT IV, type A sensor, pollution
degree 2