Fluke 1760 [75/110] General specifications

Fluke 1760 [75/110] General specifications
Specifications
General Specifications 5
5-3
General Specifications
Intrinsic uncertainty refers to reference conditions and is guaranteed for one year
Quality system developed, manufactured as per ISO 9001: 2000
Ambient conditions
Operating temp. range
Working temp. range
Storage temp. range
Reference temperature
Climatic class
Max. Voltage to earth/ Overvoltage
category
0 °C... +50 °C; 32 °F... +122 °F
-20 °C... +50 °C; -4 °F... +122 °F
-20 °C... +60 °C; -4 °F... 140 °F
23 °C ± 2 K; 74 °F ± 2 K
B2 (IEC 654-1), -20 °C... +50 °C; -4 °F +122 °F
2000 m: max. 600 V CAT IV*
power supply: 300 V CAT III
5000 m: max 600 V CAT III*
power supply: 300 V CAT II
* depending on the connected sensor
Reference conditions
Environment temp.: 23 °C ± 2 K < 60 % rH; 74 °F ± 2 K
< 60 % rH
Power frequency: 50 Hz / 60 Hz
Signal: declared input voltage Udin
Averaging: 10 minute intervals
Warmed up instrument > 3 h
Power supply: 120 V/60 Hz or 230 V/50 Hz, ± 10 %
Housing
Protection
insulated, robust plastics housing
IP40
Electrical safety
EN 61010-1 2nd edition, basic unit 300 V CAT III
entire measuring system depending on the used sensors
from 300V CATII up to 600V CATIV (1000V CATIII)
Power supply
83 to 264 V ac, 35 W, < 70 V A 45 to 65 Hz
dc: 100 to 375 V
Test voltages
Mains input - housing (earth connector)
Mains connection - measuring inputs
Measuring inputs - housing
Measuring input - measuring input
2500 V ac in the mains input circuit before fuses. Circuitry is
protected by over voltage devices behind the fuses.
2500 V ac
2500 V ac
2500 V ac