Fluke 1760 [85/110] W warning

Fluke 1760 [85/110] W warning
Options and Accessories
Accessories 6
6-7
General Characteristics
Maximum conductor size: Diameter: 15 mm.
Bus bar: 15 x 17 mm
Cable length: 2 m
Operating temperature range: 10 °C - +55 °C
Storage temperature range: 20 - +70 °C
Operating humidity: 15 % - 85 % (non-condensing)
Weight (per clamp): 220 g
Order-number: 2540445
Reference Conditions
Environment temperature range: +18 °C to +26 °C
Humidity: 20 to 75 % rh
Current: Sinusoidal waveform with 48 to 65 Hz
Distortion factor: <1 %, no DC component, stray field <40 A/m,
conductor centered within the clamp jaws
Safety Standards
IEC/EN 61010-1 2
nd
edition
IEC/EN 61010-2-032
IEC/EN 61010-031
EMC Standards
EN 61326 –1: 1997/A1: 1998.
W Warning
Utilize the clamps only on insulated conductors, max. 600 V
r.m.s. or DC to ground and frequencies below 1 kHz.
Accuracy (Typical, for 50/60 Hz)
Linearity, error in % of measured value, primary current in A: