Fluke 1760 [96/110] Flexi current sensor 100 a 500 a

Fluke 1760 [96/110] Flexi current sensor 100 a 500 a
1760
Users Manual
6-18
Flexi Current Sensor 100 A/500 A
This current probe has been designed for non intrusive, accurate AC current
measurements. Using the latest technologies (internal memory for calibration data)
provides current measurements between 1 A and 500 A. The measurement range can be
selected in the PQ Analyze software: IAC100 or IAC500.
Electrical Characteristics
Nominal current In: 100 A, 500 A AC rms
Measuring ranges: 1 A – 100 A or 5 A – 500 A AC
Peak current: 240 A, 1350 A
Overload capacity: Up to 2000 A rms
Intrinsic error: < ±1 % of mv
Linearity (10 % - 100 % of In): ±0.2 % of In.
Conductor position influence: < ±2 % of mv, distance to measuring head
>30 mm
Error due to adjacent conductor: ±2 A (Iext = 500 A, distance to head
>200 mm)
Phase error (to reference conditions): < ± 0.5 degrees
Temperature coefficient: 0.005 % of range/ °C
Safety: 600 V CAT IV, type B sensor, pollution
degree 2
General Specification
Cable length: 2 m
Length of measuring head: 45 cm (18 inch)
Operating temperature range: 10 °C - +70 °C
Storage temperature range: 20 °C - +90 °C
Operating humidity: 10 % - 80 % (non-condensing)
Weight: Approx. 0.3 kg
Order-number: 2540477
Reference Conditions
Environment temperature range: +18 °C to +26 °C
Humidity: 20 – 75 % rh
Current: Nominal value In, sinusoidal waveform, 48 –
65 Hz
Distortion factor: < 1 %. No DC component, stray field <
40 A/m, conductor centered within the Flexi
current sensor

Содержание

Скачать