Fluke 190 [125/159] Trigger and delay

Fluke 190 [125/159] Trigger and delay
Specifications
Dual Input Oscilloscope
10
115
Scope Record Mode ......... >27000 points on each input
Scope Normal Mode ............. 3000 points on each input
Scope Glitch Capture Mode ..............300 min/max pairs
on each input
Glitch Detection
5 μs to 120 s/div......... displays glitches as fast as 50 ns
Waveform Display.................A, B, A+B, A-B, A*B, A vs B
Normal, Average (2,4,8,64 x), Persistence
Time Base Accuracy...................... ±(100 ppm + 0.04 div)
Trigger and Delay
Trigger Modes........................................Automatic, Edge,
External, Video, Pulse Width
N-Cycle
Trigger Delay ................................. up to +1200 divisions
Pre Trigger View ............................. one full screen length
Max. Delay......................................................12 seconds
Automatic Connect-and-View Trigger
Source .............................................................. A, B, EXT
Slope........................................... Positive, Negative, Dual
Edge Trigger
Screen Update............Free Run, On Trigger, Single Shot
Source .............................................................. A, B, EXT
Slope...........................................Positive, Negative, Dual
Trigger Level Control Range .......................... ±4 divisions
Trigger Sensitivity A and B
DC to 5 MHz at >5 mV/div.......................... 0.5 divisions
DC to 5 MHz at 2 mV/div and 5 mV/div ...........1 division
200 MHz (FLUKE 199C, 225C) .......................1 division
250 MHz (FLUKE 199C, 225C) ..................... 2 divisions
100 MHz (FLUKE 196C, 215C) .......................1 division
150 MHz (FLUKE 196C, 215C) ..................... 2 divisions
60 MHz (FLUKE 192C)....................................1 division
100 MHz (FLUKE 19CB) ...............................2 divisions
Isolated External Trigger
Bandwidth ............................................................. 10 kHz
Modes .................................................... Automatic, Edge
Trigger Levels (DC to 10 kHz)....................120 mV, 1.2 V
Video Trigger
Standards..............................PAL, PAL+, NTSC, SECAM
Modes ......................Lines, Line Select, Field 1 or Field 2
Source............................................................................ A
Polarity ................................................. Positive, Negative
Sensitivity ....................................... 0.7 division sync level