Fluke 19xC-2x5C
Users Manual
6
The terms ‘Isolated’ or ‘Electrically floating’ are used in this
manual to indicate a measurement in which the test tool
input BNC or banana jack is connected to a voltage
different from earth ground.
The isolated input connectors have no exposed metal and
are fully insulated to protect against electrical shock.
The red and gray BNC jacks, and the red and black
4-mm banana jacks can independently be connected to a
voltage above earth ground for isolated (electrically
floating) measurements and are rated up to 1000 Vrms
CAT II and 600 Vrms CAT III above earth ground.
If Safety Features are Impaired
Use of the test tool in a manner not specified may
impair the protection provided by the equipment.
Before use, inspect the test leads for mechanical damage
and replace damaged test leads!
Whenever it is likely that safety has been impaired, the
test tool must be turned off and disconnected from the line
power. The matter should then be referred to qualified
personnel. Safety is likely to be impaired if, for example,
the test tool fails to perform the intended measurements or
shows visible damage.